NEW DELHI MATERIALS & FLUIDS TESTINGDigital report record
View report PDF

Laboratory test report

Comparative HFC fluid test results

Digital details for Anvol WG 46 and Kaizen Fire Resistant VG 46, submitted by Synergy Japan Lubricant and Chemical Company Limited.

20 results recorded · 19 pass · 1 informational

01 · Test items

Test item register

Reference comparator and candidate equivalent received for comparative evaluation.

A

Reference comparator

ANVOL WG 46

Castrol / Anvol · Pink-red HFC fluidANV-WG46-260708-A
B

Candidate equivalent

KAIZEN Fire Resistant VG 46

Kaizen / Synergy · Bright golden HFC fluidKZN-FR46-260708-B

02 · Results

Comparative test results

Methods, measured values, acceptance criteria, and conclusions transcribed from the laboratory report.

No.Test parameterMethodUnitAnvol WG 46Kaizen FR VG 46AcceptanceResult
01AppearanceVisual-Hazy red fluidClear bright goldenClear fluid, no separationPASS
02OdourSensory-Absent / mild glycolAbsent / mild glycolNo objectionable odourPASS
03Density at 15°CASTM D1298kg/L1.0701.091Report valuePASS
04Kinematic viscosity at 0°CASTM D445 / IP 71cSt405398≤405 typicalPASS
05Kinematic viscosity at 20°CASTM D445 / IP 71cSt114112108–118PASS
06Kinematic viscosity at 40°CASTM D445cSt46.0046.7545.0–48.0PASS
07Kinematic viscosity at 60°CASTM D445 / IP 71cSt24.024.323–25PASS
08Kinematic viscosity at 100°CASTM D445cSt11.2011.45Report valuePASS
09Viscosity index (VI)ASTM D2270-225249≥200 ref. / ≥240 targetPASS
10pH at 20°CpH meter-9.29.39.0–9.5PASS
11Water contentKarl Fischer / validated method% wt/wt36.035.834–38PASS
12Foaming, sequence 1IP 146 / ASTM D892mL/mL10/Nil10/Nil≤10/NilPASS
13Air release at 50°CIP 131min7.06.8≤7.0PASS
14Pour pointASTM D97 / IP 15°C-51-50≤ -50PASS
15Rust preventionIP 135A-No rustNo rustNo rustPASS
16CETOP corrosion profileCETOP RP 48H-Within reference profileWithin reference profileNo adverse metal attackPASS
17Spray ignitionNCB 570/1981-Not ignitedNot ignitedNot ignitedPASS
18Wick testNCB 570/1981-Not ignitedNot ignitedNot ignitedPASS
19Vickers 104°C vane pump wearIP 281/80mg5955≤100 interim / 500 maxPASS
20Thermal residue observationClient bench observation-Dark liquid residue after >300°C abuse heatingDarker residue after >300°C abuse heatingInformational onlyINFO

03 · Service guidance

Operating limits

Preferred bulk fluid operating range35°C–50°C

Normal recommended working range for viscosity and water retention.

Continuous upper limit≤55°C

Maintain cooling and reservoir control below this value for routine service.

Maximum bulk temperature60°C

Do not intentionally exceed; test water content and viscosity if reached.

Short excursion limit65°C

Correct immediately; inspect for water loss and additive concentration.

Localized hot surfacesAvoid >80°C–90°C

Localized heating can accelerate water evaporation.

Invalid abuse condition>300°C cup heating

Outside hydraulic service; residue formation is expected and is not a normal-use rating.